CCD Cameras¶
SES¶
Pumping SES¶
- turn cryotiger off, port 8 on pdu4 = 161.72.132.92
- turn on power for vacuum pump and valve, port 3 and 4 on pdu4.
- This is easiest with vncviewer to wifsip/archive, open a browser there and type in the IP address in the URL line.
- If asked for password, user is apc. Not the standard password.
- set temperature to 40, temperature sensor to 1 as root@ses
- cd src/sci-ccd231-ses
- ./set_temp localhost 40 40 40 40
- ./set_tsens localhost 1
- no need to wait until temperature is rather high.
- turn pump on on archive.
- ~/bin/pfeiffer-test.py –switchon
- watch pump speed.
- ~/bin/pfeiffer-test.py ;
- output like
- 001500 0011030906001500026
- 000000 0011001206000000011
- 001 0011003003001121
- first number is pump speed (1500). It will take a few minutes to reach full speed.
- open valve when pump speed is approaching 1500, on archive
- ~/bin/pfeiffer-test.py –openvalve
- speed will decrease immediately, pressure will jump up (the sensor is on the pump side of the valve)
- watch pressure and pump speed, pressure should never go up, pump speed should go back up to 1500 after some time. It helps to close the valve for a few time for a few minutes until the speed settles
- watch pressure: archive: ~/bin/ses_readpress.py
- output like
- SES: 0.009172 mbar Pump: 14.151923 mbar | T_CCD: -120.04K T_CAR: -117.44K T_x: -146.66K T_COLD: -162.60K
- first number is pressure (0.0012)
- watch pump speed: archive: ~/bin/pfeiffer-test.py
- wait until temperature is 40 for some time and pressure <= 10-5. This will take some hours to reach.
- set temperature to -120, on ses
- cd src/sci-ccd231-ses
- ./set_temp localhost -120 -200 -120 -120
- ./set_tsens localhost 0
- turn cryotiger on at pdu4
- wait as long as possible (coldhead temp at least -60)
- close valve:
- ~/bin/pfeiffer-test.py –closevalve
- turn off pump
- ~/bin/pfeiffer-test.py –switchoff
- turn off power to vacuum valve and pump on pdu4
WiFSIP¶
WiFSIP CCD¶
Voltages of the WiFSIP CCD power supply¶
- 15V: V1=15V, V2=15V
- 5V: V1=5.35V, V2=5.20V
- 30V: V1=30.5V
- 48V: V1=48V
Pumping WiFSIP CCD system¶
The vacuum in the CCD dewar needs to be maintained. Especially after extended power outages, pumping the dewar is necessary. The following steps are necessary:
Preparation¶
- AIP - The cooling system (the Cryotiger / PCC compressor) needs to be turned off, and the CCD should be heated to room temperature to avoid condensation on the CCD (it could otherwise end up being the coldest spot in the dewar)
- AIP - move the STELLA1-derotator so that the vacuum flange points downwards.
Pumping¶
- put the vacuum pump onto the table next to the telescope as seen in the picture. Connect the vacuum hose.
- Turn on the vacuum pump. Wait until it reaches about 1.E-3
- Open the Valve
- Keep pumping for as long as possible.
- Please alert AIP personal at least 2 hours before stopping. The cooling system need to be started then.
- Make note of the final pressure. Should be around 1.0 E-6
- Close Valve
- Turn off pump
- Wait for it to stop moving, then disconnect
Cleaning the WiFSIP cold head¶
If the cooling performance degrades, a possible cause can be contamination in the cold head.
This is an experimental recipe for cleaning the coldhead:
- Turn off the cooling system (Cryotiger / PCC compressor)
- Note system pressure
- Set the CCD temperature to room temperature to avoid condensation and let system warm up for several hours
- Remove gas tubes at the CCD cold head, put protective covers on valves.
- Connect valve with pressured air hoses to supply, valve with vacuum hose to return
- Open valve on return side, gas will escape audibly
- Connect N2 at the supply side, and flood the cold head with N2 gas for about 1 minute.
- Close valve on supply side, turn off N2 at bottle
- Connect vacuum pump on return side and vacuum the cold head. Reachable pressure is about 1mbar.
- Close return valve, disconnect vacuum pump and long hose, connect short hose instead (keep valve in place)
- Attach short pressured air tube on supply side (instead of long one, but keep valve in place!), fill it with ISO-Propanol / Ethanol or mixture
- Slowly open valve on supply side, re-fill ISO / Ethanol when needed - total volume is about 150ml, the solvent gets sucked into the cold head
- Let sit for at least 30 minutes
- Switch valves, connect long pressured air hose (remove short one first) at return side and connect N2
- Open supply valve (where short vacuum hose is connected), then open return (N2) valve
- Flood system with N2 for at least 30min. Carefully rotate derotator while flooding. No fluid should be visible at the short vacuum tube on the supply side.
- Close valve on return side, then close valve at supply side.
- Attach long vacuum hose on supply side (remove short one first), connect pump, turn on, open supply valve, and vacuum the cold head.
- Close valve on supply side
- Remove vacuum tube & pump, remove valve on supply side
- Attach PT-14 bottle on supply side and open it’s valve. Head will fill with PT-14.
- Remove PT-14 bottle
- Remove pressured air hose, N2 bottle, remove valve on return side
- Re-attach hoses to cold head
- Note system pressure
- Pump CCD system and re-start cooling system